No.258-L
Option : Acrlic Safery Cover
국제규격 | JIS-K7110、K7111-1、ASTM-D256、D6110、ISO-179-1:2010、180 |
---|---|
제품명 | No.258-L IMPACT TESTER (WITH REFRIGERATING MACHINE) |
기능 | This tester enables the operator to conduct the Charpy and Izod impact tests in a low temperature condition. |
제품사양 | -Temperature Range : -35 °C to 60 °C (Refrigerator) -Safety Device : Hammer Setting Sensor (Interlock System on the Upper Door), Acrylic Safety Cover -Additional Specifications : Refer to the below table -Option : Software, Electric Hammer Release Device, Torque Adjustment Device (Only for Izod) -Power Source : AC 200 V, 3-Phase, 40 A, 50/60 Hz -Dimensions/ Weight (Approx.) : W1,220 × D990 × H1,660 mm, 350 kg [Test Method : Charpy Impact Tester ] -Reference Standards : JIS-K7111, ISO-179, ASTM-D6110 -Capacity : 0.5, 1, 2, 4, 5, 7.5, 15, 25 J -Lift-up angle :150° -Impact speed (m/s)* :2.9, 3.8 (3.46) -Striking Edge Angle * : 30 ± 1° (45 ± 2°) -Striking Edge Radius (mm)* :R2 ± 0.5 (R3.17 ± 0.12) -Anvil-Edge Distance (mm)* :– -Support Length (mm)* :62 + 0.5 (101.6 ± 0.5) -Accessories : Specimen setting gauge, Block gauge, Torque wrench (Izod only) [Test Method : Izod Impact Tester ] -Reference Standards :JIS-K7110, ISO-180, ASTM-D256 -Capacity : 1, 2.75, 5.5, 11, 22 J -Lift-up angle :150° -Impact speed (m/s)* : 3.5 (3.46) -Striking Edge Angle * :75° (-) -Striking Edge Radius (mm)* :R0.8 (R0.79 ± 0.12) -Anvil-Edge Distance (mm)* :22 ± 0.2 (22 ± 0.05) -Support Length (mm)* :– -Accessories : Specimen setting gauge, Block gauge, Torque wrench (Izod only) *JIS/ISO Standard Value (ASTM Standard Value) |
No.258-L
Option : Acrlic Safery Cover